Strain gauge meter

The program Strain gauge meter is intended for various strain gauge measurements with the use of strain gauge transducers (e.g., force and torque transducers, RPM sensors, bridge and half-bridge circuits based on strain gauges, etc.) and a measurement module for strain gauges ZET 017-T.

Depending on the parameters configuration of the Strain gauge meter program, the measurements results can be represented as force, weight, displacement, torque and other parameters. Integrated generator of the measurement module can be used for power supply of the transducers.

Strain gauge meter variant of system configuration

The program Strain gauge meter is compatible only with the input channels, which are used for voltage measurements. The measurement unit by default is V, or its derivatives. In the case, if the corresponding channels are not found, the drop-down list becomes unavailable and there appears the following notification:

Strain gauge meter program - error notification

The program Strain gauge meter has an integrated signals recorder, which is used for displaying of the recorded data for the period of 60 minutes. It is also possible for the user to save the signals displayed on the graph to a file with *.dtx extension. If the program is used in multi-channel mode, then in order to display all the channels on the graph, it is necessary to enable the option Show all channels. For more convenience, the channels data is also available in chart form (the displayed parameters are as follows: max. and min. value, amplitude, average value and RMS value).

Strain gauge meter program data representation in chart form

Strain-gauge meter program: parameters configuration

Configuration of the measurement channels parameters is performed in the dialog window “Settings Strain gauge meter”, which is activated with the key “Options” in the main window of the Strain gauge meter program.

The window “Settings Strain gauge meter” is separated into several sections, which are also available for configuration: “Options”, “Strain gauge”, “Intermediate values”, “Multi-channel mode” and the section displaying the channels used for configuration.

Parameters

The section “Options” is used for configuration of general parameters of the measurement suite:

  • Power type – variants of power supply for the transducers from the integrated generator of the measurement module (constant or alternating current);
  • Indications – absolute or relative. If the relative readings are selected, it is necessary to assign two channels: measuring channel and reference channel. In the case, if absolute readings are selected, the section “Reference channel” becomes unavailable;
  • Measurement unit – the measurement unit to be displayed next to the numerical value. This parameter does not convert the measurement units and is used for informational purposes only;
  • Smoothing, ms  – the period in ms to be used for averaging of the measured values. The interval of a single measurement is 100 ms, hence, the averaging interval should be divisible by 100 ms;
  • Correction value – the value, which is not to be taken into consideration in the case, is the key “Reset in” is active. This section displays the difference between the value specified in the filed “Reset in” and the current measurements results (i.e., if we set the reset value as “0” when the current measurement value is “3”, then the correction value will be “-3”);
  • Inversion of the data – this option is used when it is necessary to obtain the results with opposite value;
  • Amplification coefficient – amplification of the measurement channel in 1, 10, 100 or 1000 times. As the amplification coefficient value is changed, the signal integral level indicator will display the ratio of the input signal to the available measurement range of the strain gauge module.

Configuration of Strain gauge meter program parameters

Meter resistance

When a load-indicating resistor is used as a sensing element, the section “Meter resistance” is used for configuration of the corresponding parameters, and the “Strain gauge” section becomes unavailable.

Strain gauge meter program parameters configuration of the Meter resistance section

The parameters also allow to assign the channel to be used for thermal compensation (i.e., to select the strain gauge to be used for temperature coefficient evaluation). The set value will be taken into consideration in the course of measurements performance.

For the load-indicating sensors, it is necessary to select the algorithm of the measurement process (i.e., to select Coefficient or Calibration file).

If there is a need to measure relative deformation, strain, or any other values having linear relation to the load-indicating sensors measurements, it is necessary to assign the Coefficient of load-indicating sensor sensitivity (the coefficient is calculated individually for each particular circuit).

In the case, if it is necessary to measure weight, force, and other physical values, which have non-linear dependence on the load-indicating sensor value, or it is difficult to calculate the sensitivity value, then it is possible to use the calibration chart for the conversion purpose. To do it, enable the option “Calibration file”. Then it is necessary to enter the load value and Strain-gauge meter value into the calibration file, or use already existing file with the calibration chart.

Upon activation of the “Calibration file” option, the user can switch over into Calibration mode by activation the corresponding option. The calibration mode implies the use of absolute values.

As the calibration mode is enabled, the indication color of the Strain gauge meter program changes (from blue to red).

The registered values correspond to the voltage level at the output of the strain gauge transducer. Thus, it is possible to conduct calibration of the sensing element by applying the reference value and recording the corresponding voltage level at the indicator of the Strain gauge meter.

Strain gauge meter operation in calibration mode
Strain gauge meter - operation in calibration mode

The section Indications of the calibration chat can be filled automatically: right-click on the reference value and select the option “Use the indicator value”.

It is also possible to save the calibration chart for further use.
Upon completion of the calibration process, uncheck the option Calibration mode and save the changes by clicking Apply key.

Strain Gauge

If a strain gauge transducer is used as a sensing element, the parameters configuration is conducted in the corresponding section “Strain gauge”. In this case, the section “Meter resistance” becomes unavailable.

It is possible to use two parameters for configuration of the strain gauge transducers:

  • Sensitivity, mV/V;
  • Limit of measurements – the maximal value of the deformation, which can be registered by the transducer.

When the strain gauge sensor is used as a primary transducer, it is possible to calculate only relative values of the strain gauge meter.

Hence, it is necessary to use a reference channel in addition to the measurement channel.

Strain gauge tab of the Strain gauge meter program

The field “Intermediate values” is used for diagnostics of the sensing element operability.

It displays the values of the measurement channel, reference channel (in the case of strain gauge transducer there is used a channel of the generator), thermal compensation channel, reference channel frequency, phase cos., intermediate result, result without reset (actual values), final result (in the case of thermal compensation channel or Reset option use, the result will be displayed with consideration of these parameters, in other cases the result will be displayed without the Reset function).

Multi-channel mode

New version of the Strain gauge meter program has a function of Multi-channel mode.

If there is a need to implement a great number of primary transducers, it becomes necessary to control each of the channels simultaneously. Previously it was possible to do it by running several copies of the “Strain gauge meter” program, which was not always convenient. Now it is possible to display several channels by using a single copy of the program and by activating the Multi-channel mode.

Parameters-configuration-of-the-Strain-gauge-meter-program-operation-in-multi-channel-mode-1
Strain gauge meter program-Selecting the number of channels to be displayed in the Multi-channel mode

The section “Multi-channel mode” allows to set the number of channels and to select a channel to be displayed for further configuration of its parameters.

To save the changes in parameters configuration, use the key “Save CFG channel” upon configuration completion for each of the channels.

Then use the “Apply” key to continue operation of the “Strain gauge meter” program.

Strain gauge meter program: prompt messages

In order to secure correct operation of the “Strain gauge meter” program, the system produces a number of prompt messages informing the user of the wrong parameters values. The developers of the program have conducted comprehensive analysis of the most frequent mistakes in the course of program operation. As a result, the program has the following notification messages:

  • Reference channel not found – the message is formed in the case if relative indications are selected, but the reference channel is not assigned;
  • Set the correct channels – the program produces this message in the case if measurement and reference channels have different sampling frequencies;
  • Reference signal level below 3% of the upper range – the message recommends to select suitable reference channel.
Strain gauge meter program - Prompt notifications - Reference channel not found
Strain gauge meter program-Prompt notifications-Wrong sampling frequencies of the reference and measurement channels
Strain gauge meter program-Prompt notifications-Low value of the reference channel in relation to the measurement range

Supported instruments

Source information of the Strain gauge meter program is represented by digital data from ZETLAB server channels. This data is formed by the output signals of the strain gauge transducers, tensile and compression gauges, bridge circuits. Measurement module ZET 017-T8 is used for digital processing of analog signals.

The Strain gauge meter program is included into Measurement software section.

Starting the program Strain Gauge meter

The Strain gauge meter program is also included into ZETLAB TENZO Software package, which is delivered together with measurement modules for strain gauges.

Attention

Failure of the programs included into the scope of ZETLAB Software package may be attributed to electrical crosstalk. In order to check operational capacity of the device, establish Ethernet connection and start the corresponding programs. Detailed description of establishing Ethernet connection is available in user manual (to enter the user manual section, start ZETLAB control panel, press F1 key and select the clause “Connection via Ethernet”).

If the device and the program operate in normal mode, you can continue using Ethernet connection.

Strain Gauge Measurement: Additional Information

ZET 7111 Digital strain gauge sensors
ZET 7111 Digital strain gauge sensors

ZET 7111 Digital Strain Gauge

Digital strain transducer  ZET 7111
Digital output CAN 2.0.
Dynamic measurements.
Digital temperature sensor ZET 7121
Digital temperature sensor ZET 7121

ZET 7121 Digital Temperature Sensor

digital output CAN 2.0
primary transducer type: resistance thermometers
CAN 2.0 interface
Digital temperature sensor ZET 7120
Digital temperature sensor ZET 7120

ZET 7120 Digital Temperature Sensor

digital output CAN 2.0
primary transducer type: thermocouple
CAN 2.0 inteface, temperature detector
Digital strain gauge sensor ZET 7110
Digital strain gauge sensor ZET 7110

ZET 7110 Digital Strain Gauge Sensor

Data interface CAN 2.0
Static measurements
Explosion-proof version
Digital meteorological sensor ZET 7023
Digital meteorological sensor ZET 7023

ZET 7023 Digital Meteorological Sensor

RS-485 interface
weather sensor for relative humidty
atmospheric pressure and temperature measuring
Hydrophone BC 311 main cover
Hydrophone BC 311 main cover

BC 311 underwater/threaded hydrophone

Measurements operation medium water, air, oil and oil products, natural gas
Operation medium temperature 0…+60 °C
Environment air temperature −60…+80 °C
ZET 7172 - main
ZET 7172 - main

Telemetering system ZET 7172

CAN 2.0 data interface
Speed of data exchange, kbps 100, 300, 1000
Offline recorder ZET-7173-main
Offline recorder ZET-7173-main

ZET 7173 offline recorder (CAN 2.0 interface)

Functionality: off-line recorder
Maximal number of addresses connected at a time (node) 32
Maximal total frequency of data recording 9,600 Hz
ZET 7175 - main
ZET 7175 - main

ZET 7175 GPS/GLONASS synchronization module

Functionality Sync Wizard
Timing accuracy 2 μs
Receiver GPS/GLONASS
ZET 7177 - main
ZET 7177 - main

ZET 7177 interface converter (CAN 2.0 → GSM)

Data transmission standard 2G, 3G
SIM-card slot Micro-SIM
Data interface CAN 2.0
Interface converter ZET 7176
Interface converter ZET 7176

ZET 7176 Interface converter (CAN 2.0 ↔ Ethernet/Wi-Fi)

Interface converter Ethernet ↔ CAN 2.0
PC connection interface Ethernet (100 Mbps)
ZETSENSOR modules connection interface CAN 2.0
Interface converter ZET 7174
Interface converter ZET 7174

ZET 7174 interface converter

Maximal sum frequency of data recording 12,000 Hz
Digital measuring module connection interface CAN 2.0
Data exchange rate 100, 300, 1,000 Kbps