FFT Spectrum analyzer of vibro-acoustic signals ZET 034
- Wide dynamic range
- Recording and processing of data received from multiple channels
- PTP synchronization
- Support of various ICP (IEPE)-standard transducers, DeltaTron / ISOTRON transducers, and CCD sensors with voltage output
- Data recording controls available at the front panel of the device
POR (price on request)
Specifications
of FFT Spectrum analyzer of vibro-acoustic signals ZET 034
Technical specifications | |
---|---|
Analog input (ADC) | |
Number of analog inputs by voltage and ICP | 4 |
ADC digit number | 24 |
Frequency ranges of the simultaneously analyzed signals | DC…20, DC…200, DC…2 000, DC…20 000 Hz |
Max. sampling frequency | 400 kHz |
Antialiasing signals filtration | up to 40 kHz |
LPF filters: Butterworth filters 120 dB/oct for sampling frequencies | 50 Hz, 500 Hz, 5 kHz, 50 kHz |
Filters with frequency correction (IEC 61672-1:2002) | А, В, С, D for 1-class precision noise meters |
Max. input voltage at unitary amplification ratio | ±10 V |
Programmable amplification ratios | 1, 10, 100, 1000 |
Input impedance | 100 kOhm |
Analog output (DAC) | |
Number of analog outputs | 1 |
Frequency range of the generated sinusoidal signal | 0,03… 20 000 Hz |
Output impedance | 50 Ohm |
Max. output current | 20 mA |
DAC digit number | 24 |
Max. output voltage value | ±10 V |
Digital input/output | |
Digital input | 8 bit |
Digital output | 8 bit |
Logic type of digital input/output | TTL |
Metrological specifications | |
Analog input (ADC) | |
Dynamic range | 110 dB |
Band-pass channels non-identity | 0,1 % |
Cross-channel phase difference | 1° per 10 kHz |
Band-pass intrinsic noise level at the maximal amplification ratio applied to the input | 1 µV |
Analog output (DAC) | |
Admissible relative error limit of the frequency set-point for the range of 0…20 000 Hz | ±0,1 % |
Admissible relative error limit of the frequency set-point for the range of 0,03…10 Hz | ±10 % |
Error limit of the set-point for output DC and AC | ±(0,2 % + 2 mV) |
Harmonic coefficient of the generated sinusoidal signal | 0,1 % |
Operational specifications | |
Dimensions | 280 × 200 × 35 mm |
Weight | 1 kg |
Type of supported transducers | ICP (IEPE), DeltaTron, ISOTRON |
Possibility of synchronization with other FFT Spectrum analyzers | yes |
Non-volatile memory volume (flash drive)* | up to 32 Gb |
Sampling frequency of data recording to the flash drive by a channel* | 50 kHz |
Time of data recording to the flash drive by all the channels at the maximal sampling frequency* | 10,8 hours |
Baud rate via HighSpeed USB 2.0 | 480 Mbps |
Baud rate via Ethernet | 100 Mbps |
* This additional option should be ordered separately