Specifications

of the ZET 7151 Digital Accelerometers

Metrological Data
Measured physical value Vibration acceleration Vibration velocity* Vibration displacement*
Frequency band 1 to 3,200 Hz
Measurement range with the BC112 sensor 1 to 2,000 m/s2 0.1 to 200 m/s 0.1 to 200 mm
Measurement error (of the measured value) ±6 %
Parameters diagnostics Power supply control, data quality control
Technical Specifications
Data refresh rate 10 Hz (RMS)
8000 Hz (raw data)
Type of connected primary transducers (PT) vibration transducer with charge output
Socket type for connecting primary transducer SMA
Data interface CAN 2.0
Transmission rate 100, 300, 1000 kbps
Operational Specifications
Dimensions 71×39×17 mm
Weight 30 g
Power supply 9 to 24 V
Current consumed at 12 V 40 mA
Reverse polarity protection yes
Operating temperature range -40 to +80 °C
Electromagnetic Compatibility, EMC
IEC 61000-4-2, ESD contact 4 kV, air 8 kV
IEC 61000-4-4, EFT power supply 0.5 kV, signal 0.25 kV
IEC 61000-4-5, Surge 500 V

* Vibration parameters correlation

Additional Information

  • Synchronization

    The interface converter synchronizes the digital sensors. This allows for observing the interaction in different points (locations of the digital sensors) of the facility.

  • Hardware Versions

    Standard: In a plastic cover (green)
    OEM: compound counting.
    Heavy Duty*: unit of standard design in an air-tight cover
    (with a transparent lid)
    *Optional

Advantages

of the ZET 7151 Digital Accelerometers

NO ADJUSTMENT IS NEEDED

The characteristics are stored in the measuring unit memory

MEASUREMENT RESULTS

The measurement results are transmitted in digital form

RELIABILITY

Stable characteristics and high reliability in operation

EX-PROOF

Ex-proof design according to CR CU 012/2011

Operating Principle

of the ZET 7151 Digital Accelerometers

The digital vibration sensor is a primary transducer (charge output accelerometer) combined with a measuring unit (ZET 7151).

The primary transducer is installed on the object to be studied and connected to the measuring unit. When vibration acts on the primary transducer, electric signal proportionate to the acting acceleration is generated at the output. The measuring unit performs analog-to-digital conversion of the sensor signal and transfers vibration values or acceleration instantaneous values via CAN interface.

The characteristics of the primary transducer are set in the device memory and measurements are performed based on those parameters. Thus, the digital sensors releases the user from setting the measuring channels and data processing.

The sensors are fixed on the measurement object while the information transmission units are mounted in electric cabinets. An advantage of the ZETSENSOR units is easy installation and uninstallation as well as indication of feed and signal transmission which, in its turn, facilitates diagnostics of the system at the facility.

Pb_free
Directive 2011/65/EU (ROHS, ROHS II)

DIGITAL ACCELEROMETERS MEET THE REQUIREMENTS OF DIRECTIVE 2011/65/EU (ROHS, ROHS II) CONCERNING LIMITATION TO USE HAZARD SUBSTANCES (PB, HG, CD, CR(VI)) IN ELECTRICAL AND ELECTRONIC EQUIPMENT; THEY ARE MANUFACTURED USING LEAD-FREE METHOD, IN COMPLIANCE WITH THE EUROPEAN COMMUNITY REQUIREMENTS

RoHS-Comliance

Application Scope

of digital accelerometer ZET 7151

  • SANITARY AND ECOLOGICAL VIBRATION CONTROL

  • WORKPLACE CERTIFICATION FOR LABOR CONDITIONS

  • SCIENTIFIC RESEARCH

  • BUILDINGS AND STRUCTURES VIBRATION CONTROL

  • MACHINES VIBRATION CHARACTERISTICS MEASUREMENT

  • PULSE PROCESS MEASUREMENTS

CONNECTING TO MEASUREMENT NETWORK

of digital accelerometer ZET 7151

Digital accelerometer ZET 7151 can be used to build distributed measurement networks. Interface converters shall be used for connecting the measurement network to the ZETLAB and ZETVIEW systems. The number of sensors connected to ZET 7174 shall not exceed 4 (an external power supply unit is required).

Additionally, ZET 7172 units are used for data transfer via a radio channel. The ZETLAB software supports connection of up to 200 gauges to a single computer.

Scheme-of-measuring-network-based-on-digital-accelerometers

Pin definition for connecting to the measurement network

Pin No. Definition
1 +(9 to 24) V
2 H line of CAN 2.0
3 L line of CAN 2.0
4 GND

BASIC CONFIGURATION

of the ZET 7151 Digital Accelerometers

DIGITAL ACCELEROMETER ZET 7151

The standard configuration package of the digital accelerometer includes:

  • Digital accelerometer ZET 7151;
  • operation documentation.

POR (price on request)

Order form - If you want to place an order or to make a request regarding technical specifications of particular product, fill out the form below:

Data protection notification

Please note that the personal data provided by you is used exclusively for the purposes of fulfilling our contractual obligations, including, but not limited to: issuing invoices, delivery documents, documentation relating to customs clearance process, etc. We do not provide your personal data to any third party except for the purposes relating to our direct contractual obligations. Upon completion of order processing and expiry of the warranty period for the Products provided by our Company, we do not store or process any of your personal data. Please note, that our Company’s Confidential policy does not imply any advertising or marketing activities with the use of your personal data (including your Company name, address, your E-mail, phone number and other information provided by you with a view to placing an Order and organization of the Delivery).

Necessary measuring equipment

supplied upon request

ZET 7174
ZET 7174
USB connection

ZET 7176
ZET 7176
Ethernet connection